Biological safety cabinets
SafeFAST Top
SafeFAST Top - Class II biological safety cabinet
Maximum personal, product and environmental protection for microbiological applications
The SafeFAST Top is a high-quality Class II biological safety cabinet, designed for laboratories with the highest demands on safety, ergonomics and hygiene. It fulfils the requirements of European standard EN 12469:2000 and is the first choice when a particularly robust, hygienic and contamination-free work area is required.
Thanks to its seamless finish made from a single piece of stainless steel with rounded corners, the SafeFAST Top offers an outstanding degree of ease of cleaning and durability - ideal for all microbiological work areas with increased hygiene requirements.
Your product benefits at a glance
- Maximum safety for user and product
- GMP-compliant stainless steel construction from a single mould
- Particularly easy to clean thanks to seamless design
- Quiet, energy-efficient operation
- Flexible expandability for your specific laboratory requirements
- Proven FASTER quality - Made in Europe
Areas of application - Versatile in research, diagnostics and production
The SafeFAST Top is suitable for all applications in which microbiological or cell-based materials are used. Typical areas of application are
- Microbiology - bacterial, mould and yeast cultures
- Virology - working with infectious viruses or samples
- Haematology - blood analysis and cell examination
- Cell culture - cultivation and manipulation of cell lines
- Molecular biology & genetics - DNA/RNA isolation, PCR preparation
- GMP-compliant production - for pharmaceutical and biotechnological processes
Safety and functional principle
The SafeFAST Top works according to the proven Class II air circulation principle:
- 70 % of the air is recirculated via an H14 HEPA main filter in the recirculation process
- 30 % of the air flows back into the room in a controlled manner via a separate H14 HEPA exhaust air filter
This air routing ensures triple protection: for the product, user and environment.
Technical features & standard equipment
- Energy-efficient: power consumption <281 W
- Quiet in operation: noise level <57 dB(A)
- High-quality stainless steel usable space made from one piece, with rounded corners
- Segmented stainless steel work surface - easy to remove and clean
- Hinged front panel for easy access and hygienic cleaning
- Multifunctional display with PLC control - all operating parameters at a glance
- 1x vacuum connection, permanently integrated
- 1x gas tap with safety solenoid valve - automatic switch-off when inactive
- 1x socket, integrated as standard
Expandable media connections & electrics
- Electrical sockets (Europe / worldwide)
- Media connections for:
- Air
- gas
- vacuum
- Compatible with international standards: DIN 12898 / DIN 12919 / DIN 3537
Optional equipment & accessories
IV hanging pole made of stainless steel
- With 6 hooks for secure suspension of infusion bags
- Ideal for pharmaceutical applications and cleanroom processes
Ergonomic footrest
- Inclined for a healthy posture
- Height-adjustable, with anti-slip coating
- Resistant to chemicals
Ergonomic laboratory chair
- Suitable for cleanliness class ISO 5 / class 100
- Height-adjustable (395-490 mm)
- Alcohol-resistant, hygienic PVC material
Base frames in two versions
- With swivel castors: 800 mm or 950 mm height, sturdy PU wheels, front with locking system
- With adjustable feet: for stationary set-up at the same height
UV-C disinfection lamp
- Automatic control via UV timer in the control panel
- UV radiation at 253.7 nm for effective decontamination
- Mounted outside the line of sight for safe working
⚠ Note: UV lamps are not a substitute for manual cleaning.
Replacement recommended every 12 months, depending on operating time.Additional customisation options (on request)
- Integrated touch monitors for documentation
- Particle counter for clean room monitoring (class A)
- Data interfaces: RS232, USB, RJ45 and many more
- Heatable work surface for IVF applications
- Special filters and pre-filter inserts
- Activated carbon filter for chemical contamination
- Potential-free contacts for external alarm forwarding
- Special solutions for GMP and FDA environments
| Model Type | SafeFAST Top 0.9 m | SafeFAST Top 1.2 m | SafeFAST Top 1.5 m | SafeFAST Top 1.8 m |
|---|---|---|---|---|
| Standards | EN12469, EN61010-1 | EN12469, EN61010-1 | EN12469, EN61010-1 | EN12469, EN61010-1 |
| General dimensions (W×H×D, mm) | 1015 × 1470 × 785 | 1320 × 1470 × 785 | 1625 × 1470 × 785 | 1930 × 1470 × 785 |
| Usable dimensions (W×H×D, mm) | 885 × 660 × 580 | 1190 × 660 × 580 | 1495 × 660 × 580 | 1800 × 660 × 580 |
| Inflow (m/s) | 0.45 | 0.45 | 0.45 | 0.45 |
| Downflow (m/s) | 0.40 | 0.40 | 0.40 | 0.40 |
| Electr. Connection | 1Ph+E - 230 V 50 hz | 1Ph+E - 230 V 50 hz | 1Ph+E - 230 V 50 hz | 1Ph+E - 230 V 50 hz |