Biological safety cabinets
SafeFAST Top

Illustrations may contain additional options subject to a charge
Illustrations may contain additional options subject to a charge
Illustrations may contain additional options subject to a charge

SafeFAST Top - Class II biological safety cabinet

Maximum personal, product and environmental protection for microbiological applications

The SafeFAST Top is a high-quality Class II biological safety cabinet, designed for laboratories with the highest demands on safety, ergonomics and hygiene. It fulfils the requirements of European standard EN 12469:2000 and is the first choice when a particularly robust, hygienic and contamination-free work area is required.

Thanks to its seamless finish made from a single piece of stainless steel with rounded corners, the SafeFAST Top offers an outstanding degree of ease of cleaning and durability - ideal for all microbiological work areas with increased hygiene requirements.

Optimised design - best performance

Your product benefits at a glance

  • Maximum safety for user and product
  • GMP-compliant stainless steel construction from a single mould
  • Particularly easy to clean thanks to seamless design
  • Quiet, energy-efficient operation
  • Flexible expandability for your specific laboratory requirements
  • Proven FASTER quality - Made in Europe
  • Areas of application – Versatile in research, diagnostics and production

    The SafeFAST Top is suitable for all applications involving microbiological or cell-based materials. Typical areas of application include:

    • Microbiology – bacterial, mould and yeast cultures
    • Virology – work with infectious viruses or samples
    • Haematology – blood analysis and cell examination
    • Cell culture – cultivation and manipulation of cell lines
    • Molecular biology & genetics – DNA/RNA isolation, PCR preparation
    • GMP-compliant manufacturing – For pharmaceutical and biotechnological processes
  • Safety and operating principles

    The SafeFAST Top operates according to the tried-and-tested Class II air circulation principle:

    • 70 per cent of the air is recirculated via a main H14 HEPA filter
    • 30 per cent of the air flows back into the room in a controlled manner via a separate H14 HEPA exhaust filter
      This airflow ensures triple protection: for the product, the user and the environment.
  • Technical Specifications & Standard Equipment
    • Energy-efficient: Power consumption <281 W
    • Quiet operation: Noise level <57 dB(A)
    • High-quality stainless steel interior made from a single piece, with rounded corners
    • Segmented stainless steel worktop – easy to remove and clean
    • Hinged front panel for easy access and hygienic cleaning
    • Multifunction display with PLC control – all operating parameters at a glance
    • 1x vacuum connection, permanently integrated
    • 1x gas tap with safety solenoid valve – automatic shut-off when not in use
    • 2x sockets, fitted as standard
  • Expandable media connections & electrical systems
    • Electrical sockets (Europe / worldwide)
    • Media connections for:
      • Air
      • Gas
      • Vacuum
    • Compatible with international standards: DIN 12898 / DIN 12919 / DIN 3537
  • Optional equipment & accessories

    Stainless steel IV hanging rail

    • With 6 hooks for the secure suspension of infusion bags
    • Ideal for pharmaceutical applications and cleanroom processes

    Ergonomic footrest

    • Angled for a healthy posture
    • Height-adjustable, with anti-slip coating
    • Chemical-resistant

    Ergonomic laboratory chair

    • Suitable for ISO Class 5 / Class 100 cleanrooms
    • Height-adjustable (395–490 mm)
    • Alcohol-resistant, hygienic PVC material

    Base frames – mobile or fixed

    • Base frame for a seated workstation height of 880 mm – The base frame can also be individually adjusted to your desired working height on request. A bespoke design is available at an additional cost.
    • With swivel castors: Mobile, sturdy polyurethane wheels, 360° swivel, with a locking system on the front wheels
    • With adjustable feet for a stationary setup, ensuring a secure, stable base with minimal vibration

    UV-C disinfection lamp

    • Automatic control via UV timer in the control panel
    • UV radiation at 253.7 nm for effective decontamination
    • Mounted out of sight for safe working

    ⚠ Note: UV lamps are not a substitute for manual cleaning.
    Replacement recommended every 12 months, depending on operating time.

  • Additional customisation options (on request)
    • Integrated touchscreen monitors for documentation
    • Particle counters for cleanroom monitoring (Class A)
    • Data interfaces: RS232, USB, RJ45 and many more
    • Heated work surface for IVF applications
    • Specialised filters and pre-filter cartridges
    • Activated carbon filters for chemical contamination
    • Potential-free contacts for external alarm signalling
    • Specialised solutions for GMP and FDA environments
Model Type SafeFAST Top 0.9 m SafeFAST Top 1.2 m SafeFAST Top 1.5 m SafeFAST Top 1.8 m
Standards EN12469, EN61010-1 EN12469, EN61010-1 EN12469, EN61010-1 EN12469, EN61010-1
General dimensions (W×H×D, mm) 1015 × 1470 × 785 1320 × 1470 × 785 1625 × 1470 × 785 1930 × 1470 × 785
Usable dimensions (W×H×D, mm) 885 × 660 × 580 1190 × 660 × 580 1495 × 660 × 580 1800 × 660 × 580
Inflow (m/s) 0.45 0.45 0.45 0.45
Downflow (m/s) 0.40 0.40 0.40 0.40
Electr. Connection 1Ph+E - 230 V 50 hz 1Ph+E - 230 V 50 hz 1Ph+E - 230 V 50 hz 1Ph+E - 230 V 50 hz